Test of a single module of the J-PET scanner based on plastic scintillators
By P. Moskal, S. Nied'zwiecki, T. Bednarski, E. Czerwiński, Ł. Kapłon, E. Kubicz, I. Moskal, M. Pawlik-Niedzwiecka, N. Sharma, M. Silarski and 20 more. 2014
152 citations recorded by the source corpus.
A Time of Flight Positron Emission Tomography scanner based on plastic scintillators is being developed at the Jagiellonian University by the J-PET collaboration. The main challenge of the conducted research lies in the elaboration of a method allowing application of plastic scintillators for the detection of low energy gamma quanta. In this paper we report on tests of a single detection module built out from the BC-420 plastic scintillator strip (with dimensions of 5 � 19 � 300 mm 3 ) read out at two ends by Hamamatsu R5320 photomultipliers. The measurements were performed using collimated beam of annihilation quanta from the 68 Ge isotope and applying the Serial Data Analyzer (Lecroy SDA6000A) which enabled sampling of signals with 50 ps intervals. The time resolution of the prototype module was established to be better than 80 ps (σ) for a single level discrimination. The spatial resolution of the determination of the hit position along the strip was determined to be about 0.93 cm (σ) for the annihilation quanta. The fractional energy resolution for the energy E deposited by the annihilation quanta via the Compton scattering amounts to σðEÞ=E � 0:044= ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi EðMeVÞ p
Subfields: Atomic and Molecular Physics, and Optics, Radiation, Radiology, Nuclear Medicine and Imaging
View Paper: https://www.semanticscholar.org/paper/6ad5fd80984774b6833f623d651caeef75d16a9d